Int J Prosthodont 31 (2018), No. 4     3. July 2018
Int J Prosthodont 31 (2018), No. 4  (03.07.2018)

Page 375-376, doi:10.11607/ijp.5781, PubMed:29624626


Effect of Software Version on the Accuracy of an Intraoral Scanning Device
Haddadi, Yasser / Bahrami, Golnosh / Isidor, Flemming
Purpose: To investigate the impact of software version on the accuracy of an intraoral scanning device.
Materials and Methods: A master tooth was scanned with a highprecision optical scanner and then 10 times with a CEREC Omnicam scanner with software versions 4.4.0 and 4.4.4. Discrepancies were measured using quality control software.
Results: Mean deviation for 4.4.0 was 36.2 ± 35 μm and for 4.4.4 was 20.7 ± 14.2 μm (P ≤ .001).
Conclusion: Software version has a significant impact on the accuracy of an intraoral scanner. It is important that researchers also publish the software version of scanners when publishing their findings.